Xprotolab reads 20.47V where it should be zero!

More
10 years 8 months ago #1108 by chosetec
I finally got around to assembling the Xprotolab kit I got a while back, and everything seems to be working, boots up to Firmware v. 2.16. When I switch to meter mode, instead of showing 0 volts on both (unconnected) channels, it goes up to a steady 20.47 Volts on both channels
This doesn't change when I connect CH1 or CH2 to GND. When I touch CH1 to +5V it reads 6.7 volts, which is also strange.
I checked all of my connections and they seem ok. I also recalibrated the offset according to the manual, and that didn't help either. Any ideas?
Brian

Please Log in or Create an account to join the conversation.

More
10 years 8 months ago #1109 by ganzziani
If you have a multimeter, check the following voltages. If any of the voltages are wrong, there might be a defective part.
  • Voltage at +5V should be between +4.75 and +5.25V
  • Voltage at -5V should be between -4.75 and -5.25V
  • Voltage at +3.3V should be between +3.2 and +3.4V
  • Voltage at pin 8 of U3 should be between +2.00 and +2.09V
TIP: If the unit is powered with more than 5.5V, the negative voltage generator would be the first component to get damaged.

Please Log in or Create an account to join the conversation.

More
10 years 8 months ago #1110 by chosetec
Thanks Gabriel,/\r\n//\r\n/I checked the voltages, and the only thing that was outside of acceptable range was the Voltage at -5V, which showed up as positive 0.55 volts on my multimeter./\r\n//\r\n/I've been powering the scope with 5.1v via USB./\r\n//\r\n/Would you have any more ideas on how to narrow down the problem? Or is it definitely the negative voltage generator?/\r\n//\r\n/Thanks for your help,/\r\n/Brian

Please Log in or Create an account to join the conversation.

More
10 years 8 months ago #1111 by chosetec
An update:/\r\n//\r\n/I re-did the solder joint on the TPS60403 legs, since on close inspection it looks like leg 5 (I think, it was the middle of 3) was incomplete. I powered up and re-calibrated, now the meter reads 0.000 on both channels, with slight fluctuations on the order of 0.001V. /\r\n//\r\n/When I touch my finger to CH1 and GND I see fluctuations up to 0.015V /\r\n//\r\n/Would you consider this noise to be within acceptable range, or do you think there could still be a bad connection or component?/\r\n//\r\n/Thanks again for all your help,/\r\n/Brian

Please Log in or Create an account to join the conversation.

More
10 years 8 months ago #1112 by chosetec
Oops, that was short-lived. The voltages are reading strangely again, around 17 volts with a lot of noise. If I order a batch of kits, would you be able to throw in a set of extra switches (I took them off and damaged them while looking for bad solder joints) and a voltage generator? /\r\n//\r\n/Brian

Please Log in or Create an account to join the conversation.