Xprotolab Specifications updated

More
15 years 4 months ago - 14 years 11 months ago #47 by ganzziani
Joshua Maserow from Silverblade Computers and Electronics CC just asked me today what happened when applying 12V on the input. I replied that the device will be fine, but the device will only measure up to 10V.
Then I got home, I applied 12V on the input, and the device measured correctly! I had miscalculated the maximum range when using the minimum gain.
The Specifications have been updated, and I also added tjhe Absolute Maximum Voltage.
Last edit: 14 years 11 months ago by ganzziani.

Please Log in or Create an account to join the conversation.

More
15 years 4 months ago #59 by ganzziani
I have been playing around checking the frequency response of the device. Unfortunetely, the XMEGA's ADC bandwaidth is not very high.
I had set up my analog filters at 320kHz, but the signals on the Xprotolab display gets attenuated to -3dB at around 200kHz.

Please Log in or Create an account to join the conversation.

More
14 years 11 months ago #209 by ganzziani
I have recalculated the input range and verified on a unit, the analog range is:
Analog Voltage Input Range: -14V to +20V

Please Log in or Create an account to join the conversation.